UL 94-2006,IEC 60695-11-4 ,IEC 60695-11-3,GB/T5169-2008
Detailed introduction to Horizontal-Vertical Flame Chamber Tester:
Burning experiment mainly refer to vertical level UL 94-2006, IEC 60695-11-4, IEC 60695-11-3, GB/T5169-2008 series standard USES the size of the provisions of the Bunsen burner (Bunsen burner) and specific gas source (methane or propane), according to certain flame height and certain ShiYan Angle to a horizontal or vertical state trial product timing was burning several times, to test the product, searing light burning duration and burning length, etc to assess its flammability and fire risk.
1, burner: inner Φ 9.5 mm + /-0.3 mm Bunsen one.
2, test Angle: 0 °, 20 °, 45 ° manual switching.
3, flame height: 20 mm ± 2 mm to 180 mm ± 10 mm is adjustable.
4, ShiYan time: 0-999.9 s + /-0.1 s adjustable.
5, YuYan time: 0-999.9 s + /-0.1 s, automatic records, wire control pause.
6, burning time: 0-999.9 s + /-0.1 s, automatic records, wire control pause.
7, burning gas: 98% methane gas or 98% propane (generally can use liquefied petroleum gas (LPG) instead of).
8, flow pressure: with double flow and pressure gauge (gas and air)
9, temperature test range: 0 ~ 1000 ℃
10, the flame temperature requirements: from 100 ℃ ± 5 ℃ went up to 700 ℃ ± 3 ℃ time in 44 ± 2 seconds and 54 seconds ± 2 seconds.
11, temperature measurement thermocouple: K type Φ 0.5 mm precision plus or minus 0.05%.
12, test process: test procedures manual control, independent convulsions
13, consult standard: UL 94 2006, IEC 60695-11-3, IEC 60695-11-4, GB/T5169.15, 16, 17, 22-2008
14, studio volume: more than 0.75 cubic (can be customized to the other size)
15, the equipment size: 1200 mm wide x 700 mm deep x 1400 mm high
16, control: the single chip microcomputer + touch screen control, the high accuracy thermometer independent type electrical control system, strong anti-jamming capability.
17, operation efficiency: test control processes to achieve a key type operation, fast and efficient test.
18, testing standard integration: according to actual test process writing test procedures.
19, automatic: automatic ignition.